Biomechanical comparison of humeral nails with different distal locking mechanisms: Insafelock nails versus conventional locking nails

Biomechanical comparison of humeral nails with different distal locking mechanisms: Insafelock nails versus conventional locking nails

1.

Bezmialem Vakif University Medical School, Department of Orthopaedics and Traumatology, Istanbul, Turkey

2.

Istanbul Medeniyet University, Department of Orthopaedics and Traumatology, Istanbul, Turkey

3.

Department of Biomechanics, Faculty of Mechanical Engineering, Istanbul Technical University, Istanbul, Turkey

AOTT 2019; 53: 490-496
DOI: 10.1016/j.aott.2019.08.003
Read: 1420 Downloads: 731 Published: 24 December 2019

Objective: The aim of this study was to compare the biomechanical resistance to rotational and axial forces of a conventional locking nail with a newly designed intramedullary humeral nail developed for humeral shaft fractures with a secure locking mechanism through the distal part of the nail.

Methods: InSafeLOCK humeral nail system (group 1, TST, Istanbul, Turkey) and Expert humeral nail system (group 2, DePuy Synthes, Bettlach, Switzerland) of the same size (9  300 mm) were examined. In total, 24 fourth-generation humerus sawbones were used in the experiment. Osteotomy was performed at the humerus shaft, and a defect was created by removing 1 cm of bone. After pre-loading 5000 cycles at a frequency of 2 Hz and a force of 50e250 N for axial loading and 5000 torsion torques between 0.5 Nm and 6.5 Nm at a 2 Hz frequency for torsional loading, the failure load values of each load were recorded. Distal interlocking was performed with an endopin in group 1, while a double cortex screw was used in group 2.

Results: All samples successfully passed the cyclic loading. The initial and final stiffness values were similar between the groups after axial loading (p ¼ 0.873 and p ¼ 0.522, respectively). The mean axial failure load values in groups 1 and 2 were 2627 ± 164 N and 7141 ± 1491 N, respectively. A significant difference was found in the axial failure load values (p ¼ 0.004). Significant differences were observed between the initial and final torsional stiffness between the two groups (p ¼ 0.004 and p ¼ 0.004, respectively). No significant difference was found in the failure load values after torsional loading (11791 ± 2055 N.mm and 16997 ± 5440 N.mm) (p ¼ 0.055).

Conclusion: These results provide a biomechanical demonstration of the adequate stability of both nails after axial and rotational loading. The reliability of the newly developed InSafeLOCK humeral nail system, which does not require fluoroscopic control and an additional incision for distal locking, supports its use in the clinic.