Acta Orthopaedica et Traumatologica Turcica

Comparison of the mechanical performance of different frame configurations of the Ilizarov fixator

AOTT 1997; 31: 343-347
Read: 587 Downloads: 368 Published: 19 April 2021
Abstract

The circular frame of Ilizarov fixator could be modified due to anatomical variations and treatment aims. But these modifications in the frame system cannot always obtain the mechanical performance of the standard frame system. A standard system has two rings in each bone fragment which are connected by four longitudinal bars on each side of the ring. In this study, mechanical performance of one standard and eight modified frame systems were compared. Each system was loaded on a Material Testing Machine with axial compression, four point bending and torsion, and overall stiffness were tested. As a result, systems that were modified with drop wires and Schanz screws 45¡ oblique to the wires on the proximal ring, provided a closer mechanical performance to standard system.

Özet

Ilizarov frame sistemi, anatomik bölgeye, yapılara ve tedavi amacına göre değişik modifikasyonlarla dü-zenlenebilmektedir. Ancak frame konfigürasyonunda yapılan değişiklikler standart frame sisteminin (her ke-mik fragmanında ikişer tam halka ve 4 rodla tesbit) mekanik performansım her zaman sağlayamamaktadır. Bu çalışmada biri standart ve sekizi standart üzerinde yapılan değişikliklerle hazırlanmış toplam dokuz fra-me sistemi mekanik performansları açısından karşılaştırılmıştır. Materyal test makinesi ile yapılan aksiyel kompresyon, dört nokta bükülme, torsiyon testleri sonucu, proksimal halkadaki tellere 45° açıyla uygulanan drop tel veya Schanz vidası ile kurulan sistemlerin standart frame sistemine yakın sıkılık sağladığı görülmüş-tür.

Files
ISSN 1017-995X EISSN 2589-1294