Acta Orthopaedica et Traumatologica Turcica

SCREWPLATE CORROSlON AND THE PROBABILITY OF SPONGIOSA TO BE CRUSHED IN COMPRESSION OSTEOSYNTHESIS

AOTT 1985; 19: 311-328
Read: 661 Downloads: 475
Abstract
The study presented below depends on biomechanical investigations and calculations. Angular autocompression plate leads to damage in spongiosa in callum femoris and trachanteric mass when an interfragmantary pressure of 200 kg/cm2 is created in a varisation osteotomy.\nWe have already observed corrosion damages and metallosis upon the removal of muitiple angular auto-compression plates. In this study attempts have been made to investigate and discuss the meohanism of corrosion damages on the basis of biomechanical conditions.\nIn order for a progressive metallosis to be prevented, metal implantations should be removed in one year, after the alte of asteotomy is united osseousiy.
Özet
Sunulan çalışma biyomekanik incelemelere, hesaplamalara dayanmaktadır. Açılı otokompresyon plağı, bir varizasyon osteotomisinde 200 kg/cm2 lik bir interfragmanter basınç yaratıldığında, kollum femoris ile trohanter kitlesinde spongioza zararlarına yol açmamaktadır.\nÇok sayılardaki açılı otokompresyon plaklarının yıllar boyu uzaklaştırılmalarında korozyon zararları ile metalloz gözlemlemiştik. Bu çalışmada korozyon zararlarının oluş mekanizması biyomekanik koşutlar temelinde incelenip tartışıldı. İlerleyen bir metailozun önlenmesi için, metal implantasyonları, osteotomi yeri ossöz olarak kaynadıktan sonra, bir yıl içinde uzaklaştırılmalıdır.
Files
ISSN 1017-995X EISSN 2589-1294