Articles

Dual plating for fixation of humeral shaft fractures: A mechanical comparison of various combinations of plate lengths

Volume 50 · Issue 4 · January 2016 Publish Date: February 7, 2020
Full Text PDF
DOI
Ahmet Karakasli
Onur Basci
Fatih Ertem
Eyad Sekik
Hasan Havitcioglu
Karakasli, A., Basci, O., Ertem, F., Sekik, E., & Havitcioglu, H. (2020). Dual plating for fixation of humeral shaft fractures: A mechanical comparison of various combinations of plate lengths. Acta Orthopaedica Et Traumatologica Turcica, 50(4), 432–436. https://doi.org/10.1016/j.aott.2016.07.002
Full Text Full Text PDF

Abstract

Objective: The role of plate configuration was found inconclusive on the biomechanical effects of the plate size and hole number for dual plate constructions in humeral shaft fractures. The purpose of this study was to test the biomechanical stability of various dual plate constructions.

Methods: Twenty-four left humeri (4th Generation Composite Humerus, Sawbones, Malmo, Sweden) € with comminuted midshaft humeral fracture were used. Four groups of plate constructs were tested: laterally fixed 8-hole locking plate and screws were combined with anteriorly locking plates containing 0, 4, 6, or 8 holes in groups I, II, III, and IV, respectively. The alterations in axial, bending, and torsional angles were recorded.

Results: There were no fixation failures during axial, bending, or torsional stiffness testing within the elastic behavior limits. Axial stiffness was highest in Group IV. Torsional stiffness, posterior-to-anterior bending stiffness, lateral-to-medial bending stiffness, and medial-to-lateral bending stiffness were lowest in Group I.

Conclusion: The similar stiffness values for the 8-to-4 hole and 8-to-6 hole plate constructions indicate that the 8-to-4 hole construction is an option in young adults, while the stiffest 8-to-8 hole combination may be an option for osteoporotic patients

Similar Articles

Article Info
Published In
Journal Acta Orthopaedica et Traumatologica Turcica
Volume / Issue Volume 50 · Issue 4 · January 2016
Pages 432-436
History
Published Online February 7, 2020
Copyright
Affiliations
1
Ahmet Karakasli
2
Onur Basci
3
Fatih Ertem
4
Eyad Sekik
5
Hasan Havitcioglu
Cite this Article
Karakasli, A., Basci, O., Ertem, F., Sekik, E., & Havitcioglu, H. (2020). Dual plating for fixation of humeral shaft fractures: A mechanical comparison of various combinations of plate lengths. Acta Orthopaedica Et Traumatologica Turcica, 50(4), 432–436. https://doi.org/10.1016/j.aott.2016.07.002
Share
Outlines