Articles

Minimally invasive plate osteosynthesis with PHILOS plate for proximal humerus fractures

Volume 51 · Issue 1 · January 2017 Publish Date: February 7, 2020
Full Text PDF
DOI
Uğur Gönç
Mesut Atabek
Kursat Teker
Altuğ Tanrıöver
Gönç, U., Atabek, M., Teker, K., & Tanrıöver, A. (2020). Minimally invasive plate osteosynthesis with PHILOS plate for proximal humerus fractures. Acta Orthopaedica Et Traumatologica Turcica, 51(1), 17–22. https://doi.org/10.1016/j.aott.2016.10.003
Full Text Full Text PDF

Abstract

Objectives: The aim of the present study was to evaluate results, including clinical and radiological outcomes and number of complications, following minimally invasive plate osteosynthesis (MIPO) of proximal humerus fractures, using the PHILOS® proximal humerus internal locking system (Synthes Holding AG, Solothurn, Switzerland).

Methods: Retrospectively evaluated were 31 patients treated with MIPO (12 male, 19 female; average age: 58.4 years). Four patients had 2-part fractures, 14 patients had 3-part fractures, and 13 patients had 4-part fractures, according to Neer classification. Healing, complications, and head-shaft angle (HSA) were radiographically evaluated. Clinical outcomes were assessed at 1-year follow-up with Constant score.

Results: Average Constant scores for fractured and normal shoulders were 73.2 ± 10.9 and 84.8 ± 5.1, respectively. Varus progression, fracture type, and age had no significant effect on functional outcome. Average postoperative and follow-up HSA's were 130.80 ± 7.70 and 128.80 ± 10.00, respectively. Significant varus progression was observed during follow-up (p &frac14; 0.01). Varus progression was more prominent in patients with postoperative HSA < 130 (p < 0.001). Inferomedial calcar screw usage, fracture type, and age had no significant effect on varus progression. Complications included 2 implant failures, 1 case of avascular necrosis (AVN), 1 primary screw cut-out, 1 axillary nerve injury, and 1 radial nerve injury (22.6% overall).

Conclusion: MIPO is a safe and effective option for the treatment of proximal humerus fractures, with good functional recovery and fewer complications, which are typically technique dependent. Reduction may be difficult, resulting in varus progression. Another disadvantage is risk of axillary nerve injury. Careful surgical technique and correct implant selection is important in the prevention of nerve injury. Level of evidence:

Level IV, Therapeutic study.

Similar Articles

Article Info
Published In
Journal Acta Orthopaedica et Traumatologica Turcica
Volume / Issue Volume 51 · Issue 1 · January 2017
Pages 17-22
History
Published Online February 7, 2020
Copyright
Affiliations
1
Uğur Gönç
2
Mesut Atabek
3
Kursat Teker
4
Altuğ Tanrıöver
Cite this Article
Gönç, U., Atabek, M., Teker, K., & Tanrıöver, A. (2020). Minimally invasive plate osteosynthesis with PHILOS plate for proximal humerus fractures. Acta Orthopaedica Et Traumatologica Turcica, 51(1), 17–22. https://doi.org/10.1016/j.aott.2016.10.003
Share
Outlines