Abstract
Objective: This study aimed to evaluate the biomechanical effects of anterior cephalomedullary nail entry-point modification, proposed to reduce distal anterior cortical contact in geriatric patients with pronounced anterolateral femoral bowing, using finite element analysis (FEA).
Methods: A three-dimensional FEA was performed using a femur model reconstructed from computed tomography data of an 80-year-old patient with pronounced anterolateral femoral bowing. A stable pertrochanteric fracture (AO/OTA 31-A1) was modeled and fixed with a proximal femoral nail antirotation (PFN-A) using three entry-point configurations: standard central entry (Model A), 5-mm anterior shift (Model B), and 10-mm anterior shift (Model C). Blade position within the femoral head and distal locking screw alignment were kept constant across the models. Nonlinear static analyses were performed under an axial load of 1400 N simulating single-leg stance. Displacement and von Mises stress distributions were evaluated for both the implant and femur.
Results: Femoral head–neck displacement was −3.19 mm in Model A, −3.23 mm in Model B, and −3.25 mm in Model C. Implant displacement was 20.20, 20.25, and 20.42 mm, respectively. Maximum implant stress, concentrated at the nail–blade junction, increased from 295.17 MPa in Model A to 333.10 MPa in Model B and 389.63 MPa in Model C. Maximum femoral stress at the bone–blade interface increased from 101.25 MPa to 138.66 MPa and 176.23 MPa, respectively. Maximum stress along the fracture line was concentrated medially and measured 68.70, 72.64, and 74.69 MPa in Models A, B, and C, respectively.
Conclusion: Anterior shifting of the cephalomedullary nail entry point had minimal effect on displacement but was associated with progressively increased stress in both the implant and femur. These findings suggest a potential biomechanical trade-off when anterior entry-point modification is used to avoid distal anterior cortical contact in patients with pronounced femoral bowing.
Cite this article as: Kismet M, Kocaoğlu H, Ertan M, Kalem M. Biomechanical effects of anterior cephalomedullary nail entry-point modification in a geriatric bowed femur: a finite element analysis. Acta Orthop Traumatol Turc.,2026; 60(6), 0907, doi: 10.5152/j.aott.2026.26907.